MYA microbalances

Balance protective cover offered in standard

 

MYA series of microbalances has been designed to meet the highest requirements for determination of mass. Measurement reliability and accuracy is ensured by system of automatic internal calibration.


Microbalances consist of two major parts (an electronic system and a precise mechanical measuring system in a separate casing). Such design eliminates the temperature influence and separates mechanical components of a microbalance from shocks and vibrations caused by users operating software.
All the elements of a microbalance are made of glass and steel which eliminates the influence of electrostatics on weighing process.

 
   
Separable Indicator
The possibility of detaching the indicator from weighing chamber significantly limits shocks and vibrations that may influence weighing process, reduces imensions and increases the ergonomics of operation.

  Infrared proximity sensors
Optional functions:
- PRINT function
- TARE function
- Opening weighing chambers
- sensor's sensitivity adjustment
  Electronic level indicator
Functions:
- ALARM function
- graphic level indicator
- programmable acceptable tilts
     
Data exchange through USB storage devices
- update balance software
- export weighing data
- export/import databases
- export/import balance settings
- exchang data between balances
  Communication interfaces
- Ethernet
- RS 232
- 2×USB
- additional display ports
   
 
Functions:
Filling
Checkweighing
Percentage
Statistics
Formulation
Air buoyancy
correction
GLP
procedures
Infrared
sensors
    
3 years warranty period
 
Download:
Product folder
Declaration
of conformity
Air buoyancy
correction -
characteristics
 
Technical data: MYA 2 MYA 5 MYA 11 MYA 21 MYA 31
Max capacity 2 g 5 g 11 g 21 g 31 g
Readability 1 µg 1 µg 1 µg 1 µg 1 µg
Tare range -2 g -5 g -11 g -21 g -31 g
Repeatability* 1 µg 2,1 µg
Linearity ±3 µg ±5 µg ±6 µg ±7 µg ±7 µg
Eccentric load deviation 3 µg 5 µg 6 µg 7 µg 7 µg
Sensitivity offset 1,5×10-6×Rt 2×10-6×Rt 3×10-6×Rt 4×10-6×Rt 5×10-6×Rt
Sensitivity temperature drift 1×10-6/°C×Rt
Sensitivity stability 1×10-6/Rok×Rt
Minimum weight (USP) 3 mg 6,3 mg
Minimum weight
(U = 1%, k = 2)
0,2 mg 0,4 mg
Pan size ø 16 mm ø 26 mm
Weighing chamber dimensions ø 90×90 mm
Stabilization time 5 s
Calibration internal (automatic)
Display 5,7'' touch screen
Interface 2×USB, RS 232, Ethernet, 4di / 4do (digital)
Rt - net weight
* repeatability is expressed as a standard deviation from 10 weighing cycles.
 
Additional equipment:
Anti-vibration table Microbalance shield
"Kafka" thermal printer "PW-WIN" computer software
Foot tare and print buttons "RAD-KEY" computer software
Standard mass "REC-FS" computer software
Professional weighing table Antistatic cable
Antistatic ionizer RS 232 cable:  scale - computer: P0108
Ambient conditions module RS 232 cable:  scale - thermal printer: P0136
Set for the determination of air RS 232 cable:  scale - "Epson/Citizen" printer: P0151
Please click entry from additional equipment list to see details 
 
©2005 RADWAG. All rights reserved.
All proprietor's rights do the materials placed in the server are in possession of RADWAG and are
not to be copied, processed and utilized in commercial purpose unless agreed with RADWAG.